A compact high resolution ion mobility spectrometer for fast trace gas analysis
文献情報
Ansgar T. Kirk, Maria Allers, Philipp Cochems, Jens Langejuergen, Stefan Zimmermann
Drift tube ion mobility spectrometers (IMS) are widely used for fast trace gas detection in air, but portable compact systems are typically very limited in their resolving power. Decreasing the initial ion packet width improves the resolution, but is generally associated with a reduced signal-to-noise-ratio (SNR) due to the lower number of ions injected into the drift region. In this paper, we present a refined theory of IMS operation which employs a combined approach for the analysis of the ion drift and the subsequent amplification to predict both the resolution and the SNR of the measured ion current peak. This theoretical analysis shows that the SNR is not a function of the initial ion packet width, meaning that compact drift tube IMS with both very high resolution and extremely low limits of detection can be designed. Based on these implications, an optimized combination of a compact drift tube with a length of just 10 cm and a transimpedance amplifier has been constructed with a resolution of 183 measured for the positive reactant ion peak (RIP+), which is sufficient to e.g. separate the RIP+ from the protonated acetone monomer, even though their drift times only differ by a factor of 1.007. Furthermore, the limits of detection (LODs) for acetone are 180 pptv within 1 s of averaging time and 580 pptv within only 100 ms.
関連文献
IF 4.616
Characterization of new N,O-dialkyl phosphoramidate-bonded stationary phases for reversed-phase HPLC – retention and selectivityIF 4.616
Integration of nanoporous membranes into microfluidic devices: electrokinetic bio-sample pre-concentrationIF 4.616
Monitoring of cellular behaviors by microcavity array-based single-cell patterningIF 4.616
Realizing nano electrospray ionization using disposable pipette tips under super atmospheric pressureIF 4.616
Raman micro-spectroscopy tracing human lymphocyte activationIF 4.616
Nanoparticle-based signal generation and amplification in microfluidic devices for bioanalysisIF 4.616
An electrochemical aptasensor for thrombin detection based on direct electrochemistry of glucose oxidase using a functionalized graphene hybrid for amplificationIF 4.616
Contents listIF 4.616
Phosphorescent sensing of Cr3+ with protein-functionalized Mn-doped ZnS quantum dotsIF 4.616
掲載誌
Analyst

Analyst publishes analytical and bioanalytical research that reports premier fundamental discoveries and inventions, and the applications of those discoveries, unconfined by traditional discipline barriers.